Optima 20

The Optica collimators are intended for integration with radiography and fluoroscopy systems. All our collimators are equipped with a long-life LED light source that minimizes any maintenance because no adjustment of the field projection lamp or the mirror is required over time.

This is a manually operated collimator with large knobs for direct and accurate adjustment of the light field. The LED can be switched on with a push-button at the front. A laser line shows the center of the beam for the correct positioning of the detector. The Optica 20 is available with and without a filter wheel for additional filtering.

Specifications

Main Features

  • Application: Stationary X-ray equipment for radiography up to 150kV tube voltage.
  • Shutters: 4 pairs of shutters in two layers.
  • Mode of operation: Manual
  • Collimation: Maximum fi eld size: 43 x 43 cm at 1m SID
  • Attenuation equivalent IEC 60601-1-3; §3.8 / §7.3 / §7.4: min. 2.0 mm Aluminium equivalent (at 75kVp, HVL 2.2mm)
  • Added filtration (optional) IEC 60601-1-3; §3.3 / §7.3 / §7.5: Models with filter unit include a manual operated filter disk with 4 positions.

    Default filter combination:

    • No added filter
    • 2 mm Al
    • 1 mm Al + 0.2 mm Cu
    • 1 mm Al + 0.1 mm Cu
  • Laser(s) (optional): Single bucky center line laser, Class 2
  • Light source: White LED (5000K). Light source can be controlled by push button, external signal (“on/off ”) and CAN-bus
  • Timers: LED timer 20 to 90 seconds, factory setting 20 seconds
    Laser(s) timer 20 to 180 seconds, factory setting 20 seconds
    The timers are software confi gurable via USB connection
  • Light field illuminance IEC 60601-2-54; §203.8.102.5: 160-220 lx at 1m from focal spot, 35x35cm. Factory setting 200 lx.
    The illuminance is software confi gurable via USB connection
  • Light field contrast IEC 60601-2-54; §203.8.102.5: 4:1 at 1m from focal spot, collimated to 35x35cm
  • Leakage radiation IEC 60601-1-3; §12.4: 0.5 mGy/hr at 1m from focal spot, with 150kVp, 4mA
  • Mounting distance (mounting plane to focal spot): 80 ± 1mm
  • Accessory rails: Two accessory slots, 178mm wide x 2 mm high
    Optional: a double height accessory rails with four slots
  • DAP meter provision: A DAP-meter of max size 177x192x20 mm can be placed inside the collimator
  • Power supply: DC model: 24VDC; 24VA
    AC/DC model: 12-45VDC; 24VA or 20-30VAC; 35VA; 50/60 Hz
  • Weight: 7.0 kg
  • Operating conditions: Ambient temperature +15°C ≤ T ≤ +40°C;
    Relative humidity 30% ≤ RH ≤ 75%; Altitude ≤ 3000m
  • Compliance: Conforms to the requirements of FDA (Class II) and the European Medical Device Directive (CE Mark, Class IIb)

Brochures


Optima 20

RocketPACS

Brochures

Quick Start Guides

Technical Support

Request Support

Call the Radincon technical support line for prompt assistance:
1300 721 734

For urgent support outside business hours, please leave a message.
This will be auto-forwarded to the on-call support staff, who will call you back as soon as possible.

If you have non-urgent support queries and requests, you can also email our support team.
Request Support