Specifications
Main Features
- Application: Stationary X-ray equipment for radiography up to 150kV tube voltage.
- Shutters: 4 pairs of shutters in two layers.
- Mode of operation: Manual
- Collimation: Maximum fi eld size: 43 x 43 cm at 1m SID
- Attenuation equivalent IEC 60601-1-3; §3.8 / §7.3 / §7.4: min. 2.0 mm Aluminium equivalent (at 75kVp, HVL 2.2mm)
- Added filtration (optional) IEC 60601-1-3; §3.3 / §7.3 / §7.5: Models with filter unit include a manual operated filter disk with 4 positions.
Default filter combination:- No added filter
- 2 mm Al
- 1 mm Al + 0.2 mm Cu
- 1 mm Al + 0.1 mm Cu
- Laser(s) (optional): Single bucky center line laser, Class 2
- Light source: White LED (5000K). Light source can be controlled by push button, external signal (“on/off ”) and CAN-bus
- Timers: LED timer 20 to 90 seconds, factory setting 20 seconds
Laser(s) timer 20 to 180 seconds, factory setting 20 seconds
The timers are software confi gurable via USB connection - Light field illuminance IEC 60601-2-54; §203.8.102.5: 160-220 lx at 1m from focal spot, 35x35cm. Factory setting 200 lx.
The illuminance is software confi gurable via USB connection - Light field contrast IEC 60601-2-54; §203.8.102.5: 4:1 at 1m from focal spot, collimated to 35x35cm
- Leakage radiation IEC 60601-1-3; §12.4: 0.5 mGy/hr at 1m from focal spot, with 150kVp, 4mA
- Mounting distance (mounting plane to focal spot): 80 ± 1mm
- Accessory rails: Two accessory slots, 178mm wide x 2 mm high
Optional: a double height accessory rails with four slots - DAP meter provision: A DAP-meter of max size 177x192x20 mm can be placed inside the collimator
- Power supply: DC model: 24VDC; 24VA
AC/DC model: 12-45VDC; 24VA or 20-30VAC; 35VA; 50/60 Hz - Weight: 7.0 kg
- Operating conditions: Ambient temperature +15°C ≤ T ≤ +40°C;
Relative humidity 30% ≤ RH ≤ 75%; Altitude ≤ 3000m - Compliance: Conforms to the requirements of FDA (Class II) and the European Medical Device Directive (CE Mark, Class IIb)